KERN & SOHN KFN-TM User Manual

Page 40

Advertising
background image

40

KFB/KFN-TM-BA_IA-e-1323

8.1 Overview non verifiable weighing systems

(contacts of circuit board [K1] not short-circuited)

Menu block
Main menu

Menu item
Submenu

Available settings / explanation

PO CHK

Weighing with
tolerance range,
see chap. 7.7

nEt H

Upper limit value „Tolerance check weighing“, input see
chap. 7.7.1

nEt LO

Lower limit value „Tolerance check weighing“, input see
chap. 7.7.1

PCS H

Upper limit value „Tolerance check counting“, input see
chap. 7.7.2

PCS L

Lower limit value „Tolerance check counting“, input see
chap. 7.7.2

BEEP

no

Acoustic signal for weighing with tolerance range
switched off

ok

Audio sound when load is within tolerance limits

nG

Audio sound when load is beyond tolerance limits

P1 REF

Zero point
settings

A2n0

Automatic zero point correction (Autozero) by changing the
display, digits selectable (0.5d, 1d, 2d, 4d)

0AUto

Zero setting range
Load range where the display after switching-on the
balance is set to zero.

Selectable 0, 2, 5, 10, 20, 30, 50, 100 %

0rAGE

Zero setting range
Load range where the display is set to zero by pressing

. Selectable 0, 2, 4, 10 , 20* , 50, 100%.

0tArE

Automatic taring „on / off“, taring range adjustable in menu
item „0Auto“.

SPEEd

Not documented

Zero

Zero point setting

P2 COM

Interface
parameter

MODE

CONT

S0 off

Continuous data output,
selectable “send zero” yes / no

S0 on

ST1

One output for stable weighing value

STC

Continuous data output of stable weighing
values

PR1

Output after pressing

PR2

Manual totalizing, see chap. 7.8.

Press

and the weighing value will be

added to the summation memory and issued.