9 service, maintenance, disposal, 1 clean, 2 service, maintenance – KERN & SOHN KFN-TM User Manual

Page 46: 3 disposal, 4 error messages

Advertising
background image

46

KFB/KFN-TM-BA_IA-e-1323

9 Service, maintenance, disposal

9.1 Clean

• Before cleaning, disconnect the appliance from the operating voltage.
• Do not use aggressive detergents (solvents or similar).

9.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized
by KERN.
Before opening, disconnect from power supply.

9.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.

9.4 Error messages

Error

message

Description

Possible causes

- - - - -

Maximum load exceeded

• Unload weighing system or reduce

preload.

- - ol - -

Err 1

Incorrect data input

• Follow format “yy:mm:dd“

Err 2

Incorrect time entry

• Follow format “hh:mm:ss“

Err 4

Zeroing range exceeded due
to switching-on balance or

pressing

(normally 4%

max)

• Object on the weighing plate
• Overload when zeroing

Err 5

Keyboard error

Err 6

Value outside the A/D changer
range

• Weighing plate not installed
• Damaged weighing cell
• Damaged electronics

Err 9

Stability display does not
appear

• Check the environmental

conditions.

Advertising