3 how to connect the platform – KERN & SOHN KFN-TM User Manual

Page 52

Advertising
background image

52

KFB/KFN-TM-BA_IA-e-1323

12.3 How to connect the platform

Disconnect the display unit from the power supply.

Solder the individual leads of the load cell cable onto the circuit board. See
diagrams below.
















PIN

Loadcell

6- conductor 4- conductor

7

EXC+

EXC+

6

SEN+

5

EXC-

EXC-

4

SEN-

3

SHIELD

SHIELD

2

SIG-

SIG-

1

SIG+

SIG+

EXC+

SEN+

SEN-

EXC-

SIG+

SIG-

SHIELD

7
6
3
2
5
4
1

LOAD
CELL