Location of seals and inscriptions, Tests – KERN & SOHN KFN-TM User Manual

Page 68

Advertising
background image

Annex page 4 of 7

Annex to Test Certificate No. DK0199-R76-11.04

Issued by DELTA

6.

Location of seals and inscriptions

Seals shall bear the verification mark of a notified body or alternative mark of the manufacturer ac-
cording to ANNEX II, section 2.3 of the Directive 2009/23/EC. The seals shall be placed so that the
enclosure can not be opened.

Location of CE mark of conformity:

The CE mark of conformity is placed on the overlay on the rear side of the device.

Inscription on the overlay:

Type, accuracy class, Temp. -10 °C / +40 °C, Certificate No. DK0199-R76-11.04.

Other inscriptions on the overlay:

Manufacturer’s name and/or logo, Part No, Supply voltage.

7.

Tests

The indicator has been tested according to EN 45501 and WELMEC 2.1 Guide for testing
of indicators.

Examination / tests

Temperature tests: 20 / 40 / -10 / 5 / 20 (tested at minimum input-voltage sensitivity)

Temperature effect on no-load indication (tested at minimum input-voltage sensitivity)

Stability of equilibrium

Repeatability

Warm-up time

Voltage variations

Short time power reductions

Electrical bursts

Electrostatic discharges

Immunity to radiated electromagnetic fields

Damp heat, steady state

Span stability

Checklist

Maximum load cell cable length and impedance of cable to load cell

Load cell interface measurements with interruptions of the sense circuit

The test item fulfilled the maximum permissible errors at all tests.