12

6.7

Linearization

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range. If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.

Carrying out linearization is restricted to specialist staff possessing

well acquainted with the workings of weighing scales.

The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter “testing instruments control”.

Observe stable environmental conditions. Stabilisation requires a

certain warm-up time.

Do not remove adjustment weight in steps 1 to 3 during linearization,

only increase it.

After successful linearisation you will have to carry out calibration;

see chapter “testing instruments control”.