KERN & SOHN KDE-T User Manual

Page 15

Advertising
background image

15

KDE-T-BA_IA-e-1210

7.5

Weighing with tare

Deposit weighing vessel. After successful standstill control press the

button. The zero display appears. The weight of the container is now internally

saved.

Weigh the material, the net weight will be indicated.

After removing the weighing container, the weight of the weighing container

appears as negative display.

The tare procedure can be repeated as many times as necessary, for example

with initial weighing of several components for a mix (add-on weighing). The limit

is reached when the total weighing range capacity is full.

To delete the tare value, remove load from weighing plate and press

.

7.6

Pre-Tare function


Setting the pre-tare function:
Using this function the weight of a tare vessel can be stored.
This value also remains saved if the balance meanwhile has been switched off and
switched on again.

Turn on weighing scale by

and wait for zero to be displayed.

Put on empty vessel and press

repeatedly until a flashing display of

PrEtArE” appears.

Press

. The current weight on the weighing pan will be saved as pre-tare

weight.

Weigh the material, the net weight will be indicated.