For condensable vapours (water vapour, solvents, ...):
☞ Do not pump vapour before pump has reached its operating

temperature and with gas ballast valve closed.

☞ Open gas ballast valve. Close gas ballast valve by turning

180°.

☞ The gas ballast valve is open if the flash on the gas ballast

cap shows to the pump.

☞ With gas ballast valve open ultimate vacuum will be reduced,

pumping speed is decreased.

☞ Use inert gas at the air inlet to avoid the formation of explo-

sive mixtures.

In case of low boiling solvents when the formation of conden-
sate is unlikely, the use of gas ballast might be unnecessary.
☞ Operating the pump without gas ballast increases the sol-

vent recovery rates at the exhaust waste vapour condenser.

gas ballast

overpressure
safety relief
device

flask at the
outlet

flask at the
inlet

In case of condensation:
Both round bottom flasks are coated with a protective layer to
prevent disintegration in case of breakage or implosion.
➨ Check liquid level in both catchpots during operation. Avoid

overflowing of the catchpots.

☞ Do not overfill the catchpots. Maximum liquid level approx.

80%, to avoid problems when removing the catchpots.

➨ Check liquid level in both catchpots regularely and drain

catchpots in time.

figure condenser
inner part

☞ Separate the condenser inner part from the vacuum installa-

tion to drain the coolant if necessary.
Therefore remove the cap on top of the condesnser. Re-
move the condenser inner part: Turn it and lift. Comply with
regulations when disposing chemicals.
When reassembling check for correct position of the seal
between condenser inner and outer part.

Removing the catchpots:
Catchpot at outlet:
➨ Remove joint clip, remove catchpot and drain condensate.

Catchpot at inlet:
➨ Admit air or inert gas (via inlet of pumping unit) to atmospheric pressure. Remove

joint clip, remove catchpot and drain condensate.

➨ Reassemble drained catchpots.
Important: Comply with regulations when disposing solvents/condensates. Reuse if
possible, purify if contaminated.

Check overpressure safety relief device at the exhaust waste vapour condenser regu-
larly, replace if necessary.
☞ Check especially for conglutination and cracks.