The controller is equipped with a short circuit proof transformer with an integrated
overload protection (no fuses).

☞ Failure of the pump (e. g. by power failure) or connected components, parts of the

supply (e. g. coolant) or change of parameters (e. g. increase of pressure in the
coolant system) must not lead to a critical dangerous situation under any circum-
stances.

Electronic equipment is never 100% fail-safe. This may lead to an indefinite status of
the equipment. Provide protective measures against misfunction and failure.
☞ Operating the pump with high or low frequency, stand still of the pump or operating

the air admittance valve must not lead to a critical dangerous situation under any
circumstances.

Ensure that in case of failure the pump and the vacuum system always will turn into a
safe status.
☞ In case of diaphragm cracks or leaks in the manifold pumped substances might be

released into the environment or into the pump housing. To reduce the risk of leaks,
ask for a diaphragm pump with additional safety diaphragm.

☞ Comply especially with notes on operation and use and maintenance.

Comply with safety regulations and safety requirements concerning the use of deep
cold cooling media (e. g. dry ice).
Attention: Deep cold media may cause severe frostbite when coming in contact with
skin.

Use only genuine spare parts and accessories.
☞ Otherwise safety and performance of the equipment as well as the electromagnetic

compatibility of the equipment might be reduced.
Possibly the CE mark or the C/US conformity becomes void if not using genuine
spare parts.

Wear parts have to be replaced regularly. In case of normal wear the lifetime of the
diaphragms and valves is > 10000 operating hours. Bearings have a typical durability
of 40000 h.

Ensure that maintenance is done only by suitably trained and supervised technicians.
Ensure that the maintenance technician is familiar with the safety procedures which
relate to the product processed by the vacuum system and that the equipment, if
necessary, is appropriately decontaminated before starting maintenance.
Comply with local and national safety regulations.
Before starting maintenance vent the system, isolate the pump and other components
from the vacuum system and the electrical supply, drain condensate and allow suffi-
cient cooling of the pump.
Before starting maintenance, wait two minutes after isolating the equipment from mains
to allow the capacitors to discharge.
Never operate the pump if covers or other parts of the pump are disassembled.
Never operate a defective or damaged pump.

Repair of the gauge head VSK 5 is not possible.

In order to comply with law (occupational, health and safety regulations, safety at work
law and regulations for environmental protection) vacuum pumps, components and
measuring instruments returned to the manufacturer can be repaired only when certain
procedures (see section “Notes on return to the factory“) are followed.