Position housing cover.
☞ Move housing cover slightly to make sure that the head co-
vers are correctly positioned.
Screw in six socket head screws fixing housing cover cross-
wise first slightly, then tighten.
☞ Do not tighten until head cover is in contact with housing,
max. torque 6 Nm.
☞ Inspect the other pump heads (side of the exhaust waste
vapour condenser) the same way.
Assemble the handle.
Obey position and orientation of the head covers and the valves
definitely.
scheme pump head with head covers and valves
valves at the outlet
(round centred
opening under
valve)
valves at the inlet
(kidney-shaped
opening beside
valve)
”side with hose connection”
”motor side”