VACUUBRAND PC 2001TE VARIO User Manual

Page 43

Advertising
background image

page 43 of 50

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 99 91 35 / 11/04/2008

Position housing cover.
☞ Move housing cover slightly to make sure that the head co-

vers are correctly positioned.

Screw in six socket head screws fixing housing cover cross-
wise first slightly, then tighten.
☞ Do not tighten until head cover is in contact with housing,

max. torque 6 Nm.

☞ Inspect the other pump heads (side of the exhaust waste

vapour condenser) the same way.

Assemble the handle.

Obey position and orientation of the head covers and the valves
definitely.

scheme pump head with head covers and valves

valves at the outlet
(round centred
opening under
valve)

valves at the inlet
(kidney-shaped
opening beside
valve)

”side with hose connection”

”motor side”

Advertising