Cleaning and assembling components – VACUUBRAND PC 2001TE VARIO User Manual

Page 45

Advertising
background image

page 45 of 50

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 99 91 35 / 11/04/2008

Round bottom flask 500 ml, coated ........................ 63 84 97

O-ring 28 x 2,5 ...................................................... 3 12 05 53
(FPM) at the spherical ground joint of the round bottom flask at
the inlet

Seal ring 19 x 24.4 ................................................... 63 72 59
(FPM) at the pressure transducer

☞ Unscrew the exhaust waste vapour condenser from the

pumping unit.

☞ Pull off old overpressure safety relief device and install new

one. Check for correct position of the PTFE-foil under the
overpressure safety relief device.

☞ Assemble exhaust waste vapour condenser to the pumping

unit.

Overpressure safety relief device ........................... 63 88 21
(at the exhaust waste vapour condenser)

Cleaning and assembling components

☞ Remove union nut at the condenser and remove hose from

the inlet of the condenser.