VACUUBRAND PC 2001TE VARIO User Manual

Page 40

Advertising
background image

page 40 of 50

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 99 91 35 / 11/04/2008

Use hex key to remove six socket head screws from pump
head and remove upper housing (housing cover, housing cover
insert and head cover).
☞ Never remove parts by using a spiky or sharp-edged tool (e.

g. screw driver), we recommend to use a rubber mallet or
compressed air (to be blown carefully into port).

Remove head cover from housing cover and check valves. Note
position of valves and remove.
☞ Replace valves if necessary.
Use petroleum ether or industrial solvent to remove deposits.
Do not inhale.

Check diaphragm for damage and replace if necessary.

Lift diaphragm carefully.
☞ Apply pressure to the clamping disc beside the diaphragm

to bring connecting rod into upper turning point position if
necessary.

☞ Never use a spiky or sharp-edged tool to lift the diaphragm.

Use diaphragm key to grip under the diaphragm to the diaphragm
support disc.
Apply pressure to the diaphragm clamping disc to bring the dia-
phragm into lower turning point position. Press diaphragm key
against diaphragm clamping disc and remove diaphragm sup-
port disc with diaphragm.
☞ If the old diaphragm is difficult to separate from the support

disc, immerse assembly in naphtha or petroleum ether. Do
not inhale!

☞ Check for washers under clamping disc. Do not mix the wash-

ers from the different heads. Make sure that the original
number is reassembled at the individual pump head.

Replacing the diaphragm:

Use Phillips screw driver to unscrew the handle at one side of
the housing cover.

It is not necessary to disassemble the exhaust waste vapour
condenser.

Use open-ended wrench to remove union nut at hose connec-
tion next to the gas ballast.
Use open ended wrench (w/f 15) to turn elbow fitting 1/4 of a
turn, reconnect hose. Do not remove the elbow fitting from the
pump head.
☞ Through reassembly a leak may result.

Advertising