Use and operation

Installing in a vacuum system

• Connection lines at the pump inlet have to be gas tight. Particles and dust must

not be aspirated, the user has to provide appropriate filters if necessary. The user

must ensure their suitability concerning gas flow, chemical resistance and safe-

ness against clogging prior to use.

• Connect an exhaust line gas tight at the pump outlet if necessary. Always dispose

of exhaust gases appropriately (e.g. into a fume hood). If there is risk of release of

dangerous or polluting fluids, install an appropriate system to catch and dispose

of those fluids.

• Reduce the transmission of vibration and prevent mechanical load due to rigid

pipelines. Insert elastic hoses or flexible elements as couplings between the pump

and rigid pipes.

Attention: Flexible elements tend to shrink when evacuated.

• The gas outlet must never be blocked. The exhaust line has always to be free

(pressureless) to ensure an unimpeded discharge of gas.

• In case of mains failure accidental venting of the system, especially with an open

gas ballast, may occur. If this may cause possible dangers provide suitable safety

measures.

• Make sure ventilation is adequate especially if the pump is installed in a housing or

if the ambient temperature is elevated. Provide external ventilation if necessary.

Avoid throttling losses by using connecting pipes with large diameter and by keeping

them as short as possible.

In case of perturbing exhaust noise connect an exhaust hose or use a silencer (see

”Accessories”).

Install outlet pipelines always falling to avoid backflow of condensate towards the

pump.

Use of a suitable valve to isolate the pump from the vacuum system is recommended

to allow the pump to warm up before pumping condensable vapours or to clean the

pump before it is switched off.

When assembling, ensure

vacuum-tightness. After assembly, check the whole sys-

tem for leaks.

Secure hose connections at the pump appropriately against accidental detaching.

The VACUUBRAND controller CVC 3000 can only be operated with components

compatible to the VACUUBRAND

VACUU•BUS system, see accessories.

The vacuum controller CVC 3000 controls VACUUBRAND diaphragm pumps NT

VARIO and VARIO-B, pumping units PC 30xx VARIO and optional coolant and vent-

ing valves. Connected components (e.g. venting valve, level sensor, external pres-

sure transducer 3000 series) are automatically identified and configurated.

Do not use more than one controller within the same VACUU•BUS system. Sev-

eral controllers in the same VACUU•BUS system will interfere with each other

and result in error messages of the connected components (pumps, valves).

NOTICE

When switching on the controller CVC 3000 for the very first time, a menu to select the language of the

controller menu is displayed. Select the desired language (e.g. ”English”) by turning the selection knob

and press to confirm. Then select the pressure unit (”mbar”, ”Torr” or ”hPa”) in the same way.

It is possible to access the language selection menu at any time by switching on the controller

while keeping the selection knob pressed.