Function vac control – VACUUBRAND PC 3012 VARIO User Manual

Page 32

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009

page 32 of 75

Vac control

Set vacuum

100 mbar

Speed

HI

Maximum Off

Delay

Off

Duration Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

00:00:00

The screen-shots show the factory-set values.

When selecting ”Graphic” the display shows a pressure vs time curve.

Preselections

Use the selection knob to select the parameters. All parameters can be altered even while operation

control is running.

Set vacuum: The ”Set vacuum” is the set point for vacuum control with pinpoint precision for NT

VARIO / VARIO-B pumps. The selection ”Turbo” leads to the optimum backing pressure for a turbo-

molecular pump.

The ”Set vacuum” is adjustable in a range of 1 - 1060 mbar or can be set to ”Turbo”.

Speed: The maximum motor speed can be preselected (to control the pumping speed).

The selection ”HI” (recommended) provides the optimum pumping speed of the pump and therefore

an optimum ultimate vacuum.

The ”Speed” is adjustable in a range of 1-100% or can be set to ”HI”.

Maximum: An upper pressure limit can be preselected. The pump switches off if the pressure limit is

exceeded, e. g. at the end of suctions or filtrations. The pressure limit is only active once the pressure

has gone below ”Maximum”.

The ”Maximum” is adjustable in a range of 1059 - 1 mbar (at the least 1 mbar higher than the ”set

vacuum”) and to ”Off”. ”Off” means that no ”Maximum” value is preset.

Delay: Delay” determines the time the coolant valve is running on after the process has been

stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to ”Off”.

Duration: Duration” determines the total process time since control start.

A preset ”Duration” (process time) is without effect, if the process is stopped due to a preset ”Maxi-

mum” before ”Duration” is reached.

The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be set to ”Off”. ”Off” means that

no endpoint of the process is defined.

Vacuum control to a preset vacuum value

• Operation on demand of a speed controlled pump (NT VARIO / VARIO-B)

The timeline in the diagram adapts automatically to the process time.

Press the selection knob twice to return to the standard display.

80 mbar

00:13:00

Function Vac control

Vac control