Tools required (metric):

- Torx screw driver TX20

- Hex key size 5

- Flat-bladed screw driver 2.5 mm

- Flat pliers

- Diaphragm key w/f 66

Cleaning and inspecting the pump heads (pumps NT VARIO / PC 300x VARIO)

The replacement of the diaphragm and the replacement of the valves can be carried out separately.

To replace the valves, remove the head covers of one side of the pump conjointly with the assembled

valve heads and fittings.

To maintain the diaphragms, the valve heads and the fittings don’t have to be disassembled. The

head covers can be removed conjointly with the assembled valve heads and fittings.

Maintain only one side of the pump at a time.

1: Cylinder pin / mark

2: Connecting rod

3: Housing

4: Washer

5: Diaphragm support disc

6: Diaphragm

7: Diaphragm clamping disc with square

head screw

8: Socket head screw

9: Blind

10: Head cover

11: Square nut

12: Fillister head screw

13: O-ring

14: Valve

15: Valve head

16: Hose nozzle

17: Connection fastener with hinged cover

18: Disc spring

19: Clamping bracket

20: Countersunk screw

21: Hose clip

22: Connection tube

10

1

9

3

2

5

6

7

8

11

18

20

13

12

14

19

16

21

17

22

15

4

View of the disassembled pump head parts