External pressure transducer VSK 3000,

capacitive, ceramic diaphragm sensor 1080-0.1 mbar ..................................................................

636657

Coolant valve VKW-B, 24 V= ....................................................................................................674220

Venting valve VBM-B / KF 16, 24 V= ...........................................................................................674217

Y-type adapter VACUU•BUS .......................................................................................................636656

Extension cable VACUU•BUS, 2m ..............................................................................................612552

Wall bushing VACUU•BUS ..........................................................................................................636153

Cable RS 232C, 9-pole, Sub-D .....................................................................................................637837

Installation set CVC 3000 (clips and screws) ..............................................................................636593

Level sensor (control of liquid level in catchpots, only for PC 3002/3003/3004 VARIO) ..............699908

Exhaust waste vapour condenser Peltronic .............................................................................699905

Silencer (with connection hose)....................................................................................................636588

Accessories

Ways to connect NT VARIO chemistry diaphragm pumps:

Small flange KF DN 16, with hose - for installation at hose nozzle..............................................677058

Small flange KF DN 16, for assembly directly at the valve head .................................................699918

(at inlet of ME 4C NT VARIO / MD 4C NT VARIO;

at outlet of ME 4C NT VARIO / MZ 2C NT VARIO)

Small flange KF DN 16, for assembly directly at the valve head .................................................699919

(at outlet of MD 4C NT VARIO)

Adapter to PTFE tubing DN 10/8 mm, for assembly directly at the valve head..........................636274

(at inlet of ME 4C NT VARIO / MD 4C NT;

at outlet of ME 4C NT VARIO / MZ 2C NT VARIO)

Adapter to PTFE tubing DN 10/8 mm, for assembly directly at the valve head..........................636275

(at inlet of MZ 2C NT VARIO; at outlet of MD 4C NT VARIO)

Elbow connecting piece (90°) for PTFE tubing DN 10/8 mm....................................................638434

T-piece for PTFE tubing DN 10/8 mm ........................................................................................638435

PTFE tubing DN 10/8 mm (sold by meter) ...................................................................................638644

Attention: Dust-laden gases, deposits and condensed solvent vapour can restrict air flow out the si-

lencer. The resultant back pressure can lead to damage of pump bearings, diaphragms, and valves.

Under those conditions, a silencer must not be used.

Cleaning the pressure transducer

Attention: Never use a spiky or sharp-edged tool to clean the pressure

transducer.

Never touch the ceramic diaphragm of the pressure transducer with hard

objects.

Fill the measurement chamber with a solvent (e. g. benzene) and allow suffi-

cient cleaning time. Observe all regulations concerning usage and disposal of

solvents!

Drain the solvent and dispose of in accordance with regulations, repeat cleaning

if necessary.

Rinse the measurement chamber several times with alcohol in order to remove

all solvent residues.

Allow the pressure transducer to dry.

Readjust the pressure transducer if necessary.

Readjustment of the controller CVC 3000

See section ”Readjustment of CVC 3000”