Assembling the connecting hoses – VACUUBRAND PC 3012 VARIO User Manual

Page 68

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009

page 68 of 75

Use an open-ended wrench w/f 14 (w/f 16 at outlet) to

slip the hoses onto the hose connectors by turning the fit-

tings.

Notes on assembling fittings and hose connections
(only MV 10C VARIO-B, PC 3010 VARIO)

The thread of the fittings at the pump heads, at the inlet and

outlet distributors as well as the hose connectors are sealed

with PTFE bond.

If the fittings or the hoses have been removed, use new PTFE

bond (included in set of seals) when reassembling.

Wind the thread of the fitting two times counterclockwise with

PTFE bond, the hose connectors three times. Ensure that the

PTFE bond ends flush with the fittings.

hose connector

thread

Note position of valve and remove.

Check valve for damage and replace if necessary. Make

sure that the valve is correctly seated.

Reassemble manifold cover and cover plate and fixate

with countersunk screws.

Replacing the valve at the distributor (outlet side)

Tighten the union nuts of the hose connections at the

pump heads with an open-ended wrench w/f 17.

Tighten union nuts first by hand and then tighten one full

turn using the open ended wrench.

Assembling the connecting hoses

Use open-ended wrench (w/f 17) to loosen the union nut

of the hose, which runs directly to the cover plate of the

distributor, at the pump head.

Use open-ended wrench (w/f 14) to turn elbow fitting 1/4 of

a turn, remove hose. Do not remove the elbow fitting from

the pump head.

Unscrew the two countersunk head screws at the manifold

cover with an Phillips screw driver and remove cover plate

and manifold cover.

Advertising
This manual is related to the following products: