To check the valves and the diaphragms use a hex key

to remove four socket-head screws from the pump head

and remove the upper part of the housing (housing cover

with housing cover insert) together with the head cover,

the valves and the O-rings.

Never remove parts by using a spiky or sharp-edged tool

(e.g. screw driver), we recommend to use a rubber mallet

or compressed air (to be blown carefully into port).

Remove the head cover carefully from the upper part of

the housing. Note the position of the valves and remove

them.

Replace valves and/or O-rings if damaged.

Use petroleum ether or industrial solvent to remove de-

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.

Outlet side

MV 10C VARIO-B / PC 3010 VARIO

MD 12C VARIO-B / PC 3012 VARIO

Inlet side

Pumps VARIO-B / Pumping units PC 301x VARIO