Description response, Operation, Command – VACUUBRAND PC 3012 VARIO User Manual

Page 46: Read commands ”cvc 3000

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999164 / 19/05/2009

page 46 of 75

0 control off

1 pump down/

determining boiling point

2 set vacuum reached/

boiling pressure found

3 current pressure below set vacuum/

automatic switch-off

0 VACUULAN

1 Pump down

2 Vac control

3 Auto mode

4 Program

5 measurement device

0: venting valve closed

1: venting valve open

0: coolant valve closed

1: coolant valve open

0: in-line valve closed

1: in-line valve open

0: pump off

1: pump on

Description

Response

XXXXXX

XXXXXXXXX

Operation

Status

process control

Fault status

Command

IN_STAT

IN_ERR

Read commands ”CVC 3000”

0: last interface command correct

1: last interface command incorrect

0: catchpot not full

1: catchpot full

0: no external fault

1: external fault

0: no fault at pressure transducer

1: fault at pressure transducer

0: no overpressure

1: overpressure

0: no fault at venting valve

1: fault at venting valve

0: no fault at coolant valve

1: fault at coolant valve

0: no fault at in-line valve

1: fault at in-line valve

0: no fault at pump

1: fault at pump