Attention: Important notes regarding the use of gas ballast

When using air rather than inert gas, risk of significant damage to equipment and/

or facilities, risk of personal injury or even loss of life exists due to the formation

of hazardous and/or explosive mixtures if air and pumped media react inside or

at the outlet of the pump.

Make sure that air/gas inlet through the gas ballast valve never leads to hazard-

ous, explosive or otherwise dangerous mixtures. If in doubt, use inert gas.

In case of condensable vapours (water vapour, solvents, ...):

Do not pump vapour until the pump has reached its oper-

ating temperature.

Open gas ballast valve. The gas ballast valve is open if the ar-

row on the gas ballast cap shows towards the marking ”GB”.

With gas ballast valve open ultimate vacuum will be re-

duced, pumping speed is decreased.

Use inert gas at the air inlet to avoid the formation of explo-

sive mixtures.

Close gas ballast valve by turning it 180°.

gas

ballast

(open)

In case of low boiling solvents when the formation of condensate is unlikely, the use

of gas ballast might be unnecessary.

Operating the pump without gas ballast increases the solvent recovery rate at the

exhaust waste vapour condenser.

If pumping condensable vapours (water vapour, solvents, ....), let the pump run with

gas ballast to reduce condensation in the pump.

Operation with silencer at the outlet: Operating the pump at high inlet pressure or

pumping dusty gases for a long time may cause clogging of the silencer. Check the

silencer regularly and replace if necessary.

In case of excess temperature, the motor is shut down by a

thermal cutout in the

winding. Attention: Reset possible only manually. Confirm error message at control-

ler by pressing key START/STOP or switch off the pump or isolate the equipment

from mains. Determine and eliminate the cause of failure. Wait approx. five minutes

before restarting the pump.

Check fan (underside the pump) regularly for dust/dirt. Clean if necessary.

A warm up period (approx. 15 min.) is required to ensure that the rated ultimate

vacuum and pumping speed are attained.