Function pump down – VACUUBRAND PC 620 NT User Manual

Page 26

Advertising
background image

page 26 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

Continuous pumping with pressure and time settings

• Operation of a vacuum pump via isolation valve

• Operation of a vacuum pump without isolation valve with VMS (Vacuum Management System,

see ”Accessories”)

Preselections

The selection of the parameters is carried out via the selection knob. All parameters are also change-

able during running operation control.

Minimum: The controller switches off the pump or closes the isolation valve when the preset value

for ”Minimum” is reached.

Minimum” is adjustable in a range of 1-1060 mbar and to ”Off”.

A preset ”Duration” (process time) is without function, if the process is stopped due to a preset ”Mini-

mum” before ”Duration” is reached.

Delay: Delay” determines the time the pump (with VMS and isolation valve) and the coolant valve

run on after stopping the process.

The ”Delay” is adjustable in a range of 1-300 minutes and to ”Off”.

Duration: Duration” determines the total process time since its start.

The process time is adjustable between 1-1440 minutes (24 h) and to ”Off”. ”Off” indicates that no

endpoint for pump down is determined.

If a ”Duration” is preset, the controller switches off the pump if the preset process time is reached,

also if a preset ”Minimum” is still not reached.

If neither ”Minimum” nor ”Duration” are preset, process control has to be stopped by pressing key

”STOP”.

The screen-shot shows the factory-set values.

Pump down

Minimum Off

Delay

Off

Duration Off

- - - - - - Graphic - - - - - -

- - - - - - Back- - - - - - - -

00:00:00

When selecting ”Graphic” the display shows a pressure over time curve.

The timeline in the diagram adapts automatically to the process time.

Press the selection knob twice to return to the standard display.

Pump down

10 mbar

00:03:50

Temporary switching from ”Pump down” to ”Vac control” (only if control is running):

Press key ”MODE”. The controller switches to function

”Vac control”, the actual vacuum is used as

set value.

The preset function of the controller does not change due to this temporary switching. When pressing

key ”STOP” the controller is in function

”Pump down” again.

Function Pump down