Troubleshooting

Fault

Possible cause

Remedy

❑ No display.

➨ Mains not plugged in?

✔ Plug in mains plug of pump-

ing unit.

➨ Device switched off?

✔ Switch on device.

➨ VACUU • BUS cable to inter-

nal power supply not plugged

in at controller?

✔ Plug in VACUU • BUS cable

at controller CVC 3000.

➨ Other causes (device defec-

tive)?

✔ Contact local distributor.

❑ Display disappears.

➨ Too much load (e.g. valves)

connected?

✔ Check current draw of the

connected devices.

➨ Short circuit at connected

valves?

✔ Replace valves.

➨ Short circuit at the RS 232

plug?

✔ Check plug and cable.

➨ Other causes (device defec-

tive)?

✔ Contact local distributor.

❑ Pressure reading incor-

rect.

➨ Device not adjusted?

✔ Readjust CVC 3000.

➨ Humidity in the measurement

chamber?

✔ Let the pressure transducer

dry, e.g. by pumping. Read-

just if necessary. Determine

and eliminate the cause for

humidity.

➨ Pressure transducer contami-

nated?

✔ See „Cleaning the pressure

transducer“.

❑ Digital pressure read-

ing is flashing, ”0.0” is

displayed.

➨ Pressure transducer not

correctly adjusted under

vacuum?

✔ Adjust CVC 3000 correctly.

❑ No digital pressure read-

ing.

➨ Pressure transducer defec-

tive?

✔ Contact local distributor.

❑ Digital pressure reading

is flashing, one blip*.

➨ Overpressure at the pressure

transducer, pressure > 1060

mbar?

✔ Release pressure immedi-

ately (risk of bursting).

❑ Warning triangle and

black valve symbol are

flashing, two blips*.

➨ External venting valve re-

moved or defective?

✔ Connect valve or replace by

a new one or reconfigure

without valve.

❑ Warning triangle and

valve symbol are flash-

ing, three blips*.

➨ NT VARIO / VARIO-B pump

and in-line valve connected?

✔ Disconnect one component,

switch controller off/on.

❑ Warning triangle and

coolant valve symbol are

flashing, four blips*.

➨ Coolant valve removed or

defective?

✔ Check connection cable of

the valve or use new valve or

reconfigure without valve.