Cleaning and inspecting the pump heads, Notice – VACUUBRAND PC 620 NT User Manual

Page 50

Advertising
background image

page 50 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

1: Cylinder pin / mark

2: Connecting rod

3: Housing

4: Washer

5: Diaphragm support disc

6: Diaphragm

7: Diaphragm clamping disc with square

head screw

8: Socket head screw

9: Blind

10: Head cover

11: Square nut

12: Fillister head screw

13: O-ring

14: Valve

15: Valve head

16:Hose nozzle

17: Connection fastener with hinged cover

18: Disc spring

19: Clamping bracket

20: Countersunk screw

21: Hose clip

22: Connection tube

Cleaning and inspecting the pump heads

The replacement of the diaphragm and the replacement of the valves can be carried out separately.

To replace the valves, the head cover has not to be removed.

To maintain the diaphragms, the valve heads and the fittings don’t have to be disassembled. The

head covers can be removed conjointly with the assembled valve heads and fittings.

View of the disassembled pump head parts

(fig.: MD 4C NT)

Replacing diaphragms and valves is described for a pump

MZ 2C NT (PC 510/511/520 NT).

Perform replacing diaphragms and valves at pumping units PC

610/611/620 NT analogously (see also ahead).

NOTICE

10

1

9

3

2

5

6

7

8

11

18

20

13

12

14

19

16

21

17

22

15