Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

• Comply with

maximum permissible pressures at inlet and outlet and pressure

differences, see section ”Technical data”. Do not operate the pump with overpres-

sure at the inlet.

• Check that mains voltage and current conform with the equipment (see rating

plate).

• Connect pipes gas tight at inlet and outlet of the pump.

Attention: Flexible elements tend to shrink when evacuated.

Provide a firm level platform for the equipment and check that the system to be

evacuated is mechanically stable and that all fittings and hose connections are se-

cure. Ensure a stable position of the pump without any mechanical contact except of

the pump feet. Comply with all applicable safety regulations.

Keep a distance of minimum 20 cm between fan and ambient parts (e.g. housing,

walls, ...). Check fan regularly for dust/dirt, clean if necessary to avoid a cutback of

ventilation.

If the equipment is brought from cold environment into a room for operation, allow

the equipment to warm up (pay attention to water condensation on cold surfaces).

The diameter of the inlet and outlet pipeline should be at least as large as the diam-

eter of the pump connection pipelines.

Comply with all

applicable and relevant safety requirements (regulations and guide-

lines),

implement the required actions and adopt suitable safety measures.

Ambient conditions

To the best of our knowledge the equipment is in compliance with the requirements

of the applicable EC-directives and harmonized standards (see ”Declaration of con-

formity”) with regard to design, type and model. Directive IEC 1010 (EN 61010) gives

in detail conditions under which the equipment can be operated safely (see also IP

degree of protection).

Adopt suitable measures in case of differences, e. g. using the equipment outdoors,

installation in altitudes of more than 1000 m above mean sea level, conductive pol-

lution or bedewing.

Pay attention to the

permissible maximum ambient and gas inlet temperatures

(see ”Technical data”).

Operating conditions

The pumps have no approval for operation in or for pumping of potentially

explosive atmospheres.

The pumps are

not suitable to pump

-

unstable substances and substances which react explosively under

impact (mechanical stress) and/or when being exposed to elevated

temperatures without air,

-

self inflammable substances,

- substances which are

inflammable without air and

-

explosive substances.

Secure coolant hoses at the hose nozzles (e.g. with hose clip) to prevent their

accidental slipping.

Check the

overpressure safety relief device at the exhaust waste vapour con-

denser in appropriate intervals.

Avoid overpressure of more than 0.2 bar in case inert gas is connected to the

pump, the gas ballast connection or to a venting valve.

NOTICE

NOTICE