Operation

Command Parameter Description

function

OUT_MODE X

1: continuous pumping

2: vacuum control without automatic

3: vacuum control with automatic

set vacuum

OUT_SP_1

XXXX

unit according to preselection

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

set vacuum

OUT_SP_V

XXXX

unit according to preselection

with venting*

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

selected frequency

OUT_SP_2

XX.X

motor frequency in Hz (01.0 to 60.0 in steps

of 0.5 Hz or 99.9 for “HI”)

pressure for restart

OUT_SP_3

XXXX

unit according to preselection

(VACUU•LAN)

(see OUT_SP_1)

delay

OUT_SP_4

XX:XX

hh:mm (hours:minutes)

vacuum for automatic OUT_SP_5 XXXX

unit according to preselection

switching off

(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)

time for automatic

OUT_SP_6

XX:XX

hh:mm (hours:minutes)

switching off (VACUU•LAN)

starting process control START

stopping process control STOP

X

1: termination of process control

2: termination of process control and storage of

the current pressure as new set point

Remote operation**

REMOTE

X

0: set controller to local operation

1:

set

controller

to

remote

operation

* Pressure setting with venting is only possible in function ”Vacuum control” if an venting valve is

connected and configurated and vacuum control is started. The venting valve opens automatically

if the actual pressure is at least 10 mbar below the preset pressure. Automatic venting becomes

inactive if vacuum control is stopped (STOP or VENT), setting a pressure value using the command

OUT_SP_1 or if the function is changed. Activate the command OUT_SP_V again if necessary.

** If remote operation is selected or deselected, the user has to ensure that no dangerous status of

the system due to the change of the mode of operation can occur and to provide appropriate safety

measures, especially if selecting remote operation interferes with a locally operated active process.

Write commands CVC 2000

2: RxD

5: Mass

9: +5V (Bluetooth)

3:

TxD

7:

RTS

4: DTR

8: CTS

Connection plug arrangement

driving venting valve

OUT_VENT

X

0: close venting valve (not automatically)

1: open venting valve

(process control stopped)

30: sensitivity: low

31: sensitivity: normal

32: sensitivity: high

(optional)