Function configuration – VACUUBRAND PC 620 NT User Manual

Page 34

Advertising
background image

page 34 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

In the menu ”Configuration” the device parameters are preselected.

Adjustment: Adjustment of the pressure transducer under vacuum and/or at atmospheric pressure,

see also section ”Readjustment of CVC 3000”.

Adjustment at atmospheric pressure is carried out at an absolute vacuum value between 1060 - 700

mbar and under vacuum at an absolute vacuum value between 0 - 20 mbar. In the range between 20

and 700 mbar no adjustment is possible; ---- mbar is displayed.

RS-232: Configuration of the interface, setting of parameters and commands, see section ”Inter-

face”.

Baud rate is selectable on 19200, 9600, 4800 oder 2400, parity on ”8-N-1”, ”7-O-1” oder ”7-E-1”,

Handshake on ”no”, ”Xon-Xoff” or ”RTS-CTS” and remote on ”On” or ”Off”.

Sensors: Selection of the pressure transducer to be controlled (maximum 4 external sensors).

Display: Selection of the device parameters ”Brightness” between 0 - 100%, ”Contrast” between

0 - 100%, ”Sound” ”On” or ”Off”, ”Units” ”mbar”, ”hPa” or ”Torr”, ”Language” ”German”, ”English”,

”French”, ”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”, ”Russian”, ”Polish”,

”Dutch”, ”Japanese”

Autostart: If ”Autostart” is set to ”On” the controller restarts a running process automatically after a

mains failure. If this is undesired, set ”Autostart” to ”Off”.

Attention: If ”Autostart” is preselected, the process starts immediately after power failure without

pressing any further key. It is the user’s responsibility to ensure that no dangerous status of the sys-

tem due to the automatic start-up can occur and to provide appropriate safety measures. If neces-

sary, the user has to check

prior to using the equipment if the option ”Autostart” is enabled.

Defaults: If ”Defaults” is set to ”Load”, the controller is reset to factory set values. All stored programs

are deleted.

The screen-shot shows the factory-set values.

Preselections

The selection of the parameters is carried out via the selection knob.

Function Configuration

Configuration

Adjustment

1013 mbar

RS-232...

Sensors...

Display...

Autostart Off

Defaults

Cancel

- - - - - - - - - Back - - - - - - - - - -