Example for use – VACUUBRAND PC 620 NT User Manual

Page 30

Advertising
background image

page 30 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

Program

N0

hh:mm:ss

Vac Vent. Step

01 00:00:00 ATM

02 00:05:00 10

03 00:15:00 10

04 00:01:00 500

05 00:09:00 500

06 00:10:00 5

07 00:20:00 5

08 00:01:00 ATM

09 00:00:00

10 00:00:00

- - - - - - - - - - - - - Back - - - - - - - - - - - - -

01:01:00

Program step 1 should be always a definite initial state, here atmospheric pressure. To reach this state

definitely set a tickmark at venting and ”Step” (press selection knob).

Step 2 causes pumping down to 10 mbar within 5 minutes (linear ramp) as ”Step” is not set.

Step 3 keeps the vacuum (10 mbar) constant for 15 minutes.

Step 4 vents within 1 minute to 500 mbar.

Step 5 keeps 500 mbar for 9 minutes.

Step 6 pumps down to 5 mbar within 10 minutes.

Step 7 keeps the vacuum (5 mbar) constant for 20 minutes.

Step 8 vents as fast as possible to atmospheric pressure and switches off the control after one minute.

Vacuum pump with isolation valve and/or Vacuum-Management-System Module A: Pumping

down with intermediate venting

Example for use

Once the program is finished, the clock symbol starts to flash. Confirm the end of the program by press-

ing START/STOP (clock symbol will disappear).

Attention: If ”Autostart” is set to On, the program will start again (time will be reset to 00:00:00)

after a power failure or after switching the controller off/on. Only if the end of the program (clock symbol

flashing) has been confirmed by pressing START/STOP, the program will not start again.