Maintenance and repair, Notice – VACUUBRAND MZ 8C User Manual

Page 19

Advertising
background image

page 19 of 52

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Three-phase motors are equipped with a circuit- breaker

which isolates all poles in case of an overload.

Note: Only manual reset is possible. Switch off the pump

and disconnect from the power source. Identify and elimi-

nate the cause of failure. Allow the pump to cool down suf-

ficiently before restart. Wait approx. five minutes before

restarting the pump.

Maintenance and repair

In order to comply with laws (occupational, health and

safety regulations, safety at work law and regulations for

environmental protection) vacuum pumps, components

and measuring instruments can only be returned when

certain procedures (see section „Notes on return to the

factory“, pg. 47) are followed.

Take advantage of our service seminars, which put special

focus on the maintenance and repair of vacuum pumps.

For details and for the online ”Instructions for repair” man-

ual see www.vacuubrand.com.

In normal use, the lifetime of the diaphragms and valves

is typically 15,000 operating hours. Bearings have a typi-

cal durability of 40000 h. Motor capacitors have a typi-

cal durability in the range of 10000 to 40000 h depending

strongly on operation conditions including ambient tem-

perature, humidity or load.

NOTICE

Advertising
This manual is related to the following products: