Technical data

General technical data valid for all pumps

Maximum permissible inlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum permissible outlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum pressure difference between

inlet and outlet

psi

(bar)

16

(1.1)

Maximum permissible pressure

(absolute) at gas ballast valve

psi

(bar)

17.5

(1.2)

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 50 to 104

(-10 to +60 / +10 to +40)

Permissible relative atmospheric mois-

ture during operation (no condensation)

%

30 to 85

Maximum permissible installation alti-

tude above mean sea level

ft

(m)

6500

(2000)

Rated motor power

hp (kW)

0.52 (0.39)

No-load speed 50/60 Hz

rpm

1500 / 1800

Maximum permissible range

of supply voltage

Attention: Observe specifications of

rating plate!

100 V~ ±10% 50/60 Hz

120 V~ +5%/-10% 60 Hz

230 V~ ±10% 50/60 Hz

400 V 3~ ±10% 50 Hz

Motor protection

single-phase

three-phase

thermal cutout, manual reset

in case of an overload, all

poles are isolated

Degree of protection IEC 529

IP 20

Inlet

small flange KF DN 25

Outlet

hose nozzle for tubing I.D. 3/8”

(hose nozzle DN 10 mm)

A-weighted emission sound pressure

level* (uncertainty K

pA

: 3 dB(A))

dB(A)

52

* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and ulti-

mate vacuum with exhaust tube at outlet.

We reserve the right for technical modification without prior notice!