Shutdown & storage, Notice – VACUUBRAND MZ 8C User Manual

Page 32

Advertising
background image

page 32 of 52

- With gas ballast valve open, the ultimate vacuum will

be reduced.

- Use inert gas for gas ballast to avoid the formation of

explosive mixtures. Attention: maximum supply pres-

sure of inert gas: 17.5 psi (1.2 bar) absolute.

- Close the gas ballast valve by turning the cap 180°.

In case of low boiling solvents (when the formation of con-

densate is unlikely), the use of gas ballast might be unnec-

essary. Operating the pump without gas ballast increases

the solvent recovery rate at the exhaust waste vapor con-

denser (see „Accessories“, pg. 33).

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

Has the pump been exposed to media which may damage

the pump materials or form

deposits?

Check and clean pump heads if necessary.

Long-term:

Take measures as described above regarding short-term

shutdown.

Separate the pump from the application.

Close the gas ballast valve.

Close inlet and outlet ports (e.g., with transport caps).

Store the pump under dry conditions.

NOTICE