Important notes regarding the use of gas ballast, Notice – VACUUBRAND MZ 8C User Manual

Page 31

Advertising
background image

page 31 of 52

eration. This may reduce or block the fans’ air supply.

Check fan regularly for dust/dirt. Clean fan guard grill if

necessary to avoid a reduction of ventilation.

A warm up period (approximately 15 min.) is required to

ensure that the rated ultimate vacuum and pumping speed

are attained.

Important notes regarding the use of gas bal-

last

Gas ballast is a continuous purge to keep the pump’s in-

terior as clean as possible and to reduce the possibility of

condensation inside the pump.

Air and pumped media might react inside the pump or

at the outlet of the pump and form hazardous or explo-

sive mixtures, when you use air rather than inert gas

for the gas ballast. This constitutes a risk of significant

damage to equipment and/or facilities, a risk of per-

sonal injury or even loss of life.

+

Make sure that air/gas intake through the gas ballast

valve can never lead to hazardous, explosive or other-

wise dangerous mixtures. If in doubt, use inert gas.

To reduce condensation in the pump, do not pump vapor

before the pump has reached its operating temperature.

Open the gas ballast valve when pumping condensable

vapors. Turn gas ballast cap to open valve.

For

condensable vapors (water vapor, sol-

vents, etc.):

-

ME 16C / MZ 8C / MD 8C: The gas ballast

valve is open if the arrow on the gas ballast

cap is pointing towards the side of the pump

with the rating plate.

MD 12C / MV 10C: The gas ballast valve is

open if the arrow on the gas ballast cap is

pointing away from the pump‘s inlet.