Check the performance by measuring the vacuum at the inlet port of the

individual pump head:

Use a suitable vacuum gauge (e.g., DVR 2, cat. no.: 682902), being sure

that it is correctly calibrated. Measure the vacuum at the inlet port (marked

IN). A vacuum of less than 120 mbar should be achieved.
+

If the reading is higher, recheck the pump chamber and make sure

that the valves and the diaphragms are correctly seated (diaphragms

concentric with bore).

Individual performance check of a pump head:

Replace diaphragms and valves of all eight pump heads

in the same way.