Gas inlet temperatures, Wetted parts, Abbreviations – VACUUBRAND MZ 8C User Manual

Page 25

Advertising
background image

page 25 of 52

Gas inlet temperatures

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

Wetted parts

Components

Wetted materials

Outlet

ETFE

Inlet

Stainless steel

Diaphragm clamping disc

ETFE carbon fiber reinforced

Diaphragm

PTFE

Valves

FFKM

Valves (ME 16C)

PTFE

Head cover

ETFE carbon fiber reinforced

Housing cover insert

PTFE carbon reinforced

O-ring (only MV 10C)

FPM

Fittings

ETFE

Tubing

PTFE

Abbreviations

ETFE: Ethylene/Tetrafluoroethylene

FFKM: Perfluoro elastomer

FPM: Fluoroelastomer

PTFE: Polytetrafluoroethylene

We reserve the right for technical modification without prior notice!