VACUUBRAND MV 2 VARIO User Manual

Page 15

Advertising
background image

page 15 of 41

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999035 / 15/12/2009

☞ Condensate and deposits in the pressure transducer affect the measuring result.

☞ In case of deposits, aggressive or condensable media, install a gas washing bottle before the pressure

transducer if necessary.

☞ In order to avoid malfunction it is important to position the pressure transducer in the vacuum line so as

to avoid flow of condensate towards the pressure transducer.

☞ Clean pressure transducer if necessary, see section ”maintenance”.

gauge head
VSK 5

Shutdown:

Short-term:
Has the pump been exposed to condensate?
➨ Allow the pump to continue to run at atmospheric pressure for a few minutes (con-

tinuous pumping, setting ”HI”).

Has the pump been exposed to media which may damage the pump materials or forms
deposits?
➨ Check and clean pump heads if necessary.

Long-term:
➨ Take measures as described in section short-term shutdown.

➨ Separate pump from the apparatus.

➨ Close inlet and outlet port (e. g. with transport caps).

➨ Store the pump in dry conditions.

Max. permitted pressure at the pressure transducer: 1.5 bar (ab-
solute).
Attention: Risk of bursting!
☞ The display flashes at a pressure higher than 1100 mbar.

☞ Comply with max. permitted pressure.

☞ Inside a vacuum system where evaporation occurs, e. g.

rotary evaporator, the vacuum is not uniform, e. g. a cold
condenser acts as pump or the vacuum in the pipeline is
lower than in the system. Therefore carefully choose posi-
tion where to connect the gauge head.

➨ Connect the cable of the gauge head to the female connec-

tor at the rear side of the vacuum controller.

➨ Connect the gauge head to the vacuum system by using a

small flange connection or a hose connection.

☞ The device is adjusted together with the gauge head at the

factory. If the gauge head is replaced a readjustment is rec-
ommended.

➨ Connect cable from the frequency converter of the pump to

the controller (rear side): trip line and power supply of the
frequency converter.

female connector to
connect gauge head
VSK 5