VACUUBRAND MV 2 VARIO User Manual

Page 22

Advertising
background image

page 22 of 41

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999035 / 15/12/2009

Manual determination of the set point p

➨ Setting the value for the speed by using key ▲ or ▼ (fac-

tory-set: ”HI”).

☞ To activate set mode: Press key shortly. The current speed

is displayed for one second.

☞ With a second tip within one second or continuous pressing:

☞ Set speed. The new set point is stored when the key is re-

leased.

To start process control:
➨ Start: Press key START and immediately key p▼. Keep key

p

▼ pressed. Pumping down starts with preselected speed.

☞ Supervise process permanently. When sufficient evapora-

tion appears, release key p

▼.

☞ The actual pressure is stored as new set point.

☞ Pressure control starts.

Adjusting the process pressure by pumping down:
➨ Press key p▼.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous pressing:

Pumping down with preset speed, the actual pressure is dis-
played and stored as new set point p when the key is re-
leased.

☞ The achieved pressure is stored as new set point and ad-

justed automatically.

67

mbar

p

520

mbar

p

Adjusting of set point p during pressure control

45.0

START

STOP

p

Adjusting the process pressure upwards
➨ Press key p▲.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous pressing:

The air admittance valve is opened if connected or pressure
increase due to system parameters, the current pressure is
displayed.

☞ The achieved pressure is stored as new set point when the

key is released.

MODE

Temporary switching to setting continuous pumping

☞ Switching if process control is inactive:

➨ Keep key MODE pressed while pressing key ▲.

Advertising