Read commands – VACUUBRAND MV 2 VARIO User Manual

Page 26

Advertising
background image

page 26 of 41

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999035 / 15/12/2009

Read commands

Function

Command Response

Description

actual pressure

IN_PV_1

XXXX mbar or

unit according to preselections

XXXX Torr or
XXXX hPa

preset pressure

IN_SP_1

XXXX mbar or

unit according to preselections

(set point p)

XXXX Torr or
XXXX hPa

actual pumping speed

IN_PV_2

XX.X Hz

device set

IN_CFG

X0X0X

preselections

0: remote operation off
1: remote operation on
0: no air admittance valve
1: air admittance valve
1: continuous pumping
2: pressure control
4: TURBO•MODE

malfunction

IN_ERR

XXXX

1: last command to interface incorrect
1: malfunction pressure transducer
1: overpressure
1: fault frequency converter

status of

IN_STAT

0XXX

process control

10: continuous pumping: inactive
11:

active

20: pressure control:

inactive

21:

actual pressure >
selected pressure

22:

actual pressure =
selected pressure (+/- 1mbar)

23:

actual pressure <
selected pressure

40: TURBO•MODE:inactive
41:

actual pressure >
selected pressure

42:

actual pressure < /=
selected pressure

0: air admittance valve not driven (closed)
1: air admittance valve driven (open)