Setting the process parameter

Changing the backing pressure:
➨ Press p▲ or p▼.

☞ Set p and the backing pressure are displayed for approx. 1s.

☞ With a second tip within one second or continuous pressing:

➨ Set the backing pressure by pressing key p▲ or p▼.

☞ The value for the backing pressure is stored as the new set

point when the key is released.

➨ Press p▲ or p▼.

☞ The backing pressure (factory set: ”Lo” ) is displayed.

☞ If ”Lo” is selected, an automatic control program is started

to determine the optimum motor speed to reach the best
ultimate vacuum of the diaphragm molecular pump.

p

p

Note: If the diaphragm pump is used as backing pump of a turbo molecular pump, it is recommended to
operate the diaphragm pump in basic mode ”TURBO•MODE” and setting ”Lo”.