Interface parameters, B.96, P.no – VACUUBRAND MV 2 VARIO User Manual

Page 25: H.no, R.no

Advertising
background image

page 25 of 41

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999035 / 15/12/2009

Interface parameters

The controller CVC 2000

II

is equipped with a serial interface at the rear side of the housing (RS 232C, nine-

pole Sub-D-plug).
☞ Respectively plug-into or remove the cable from the interface (cable RS 232C, nine-pole Sub-D) only if

the equipment is switched off.

☞ The interface is not electrically isolated from the measuring circuit.

☞ For optimal electromagnetic compatibility assemble an interface filter (cat. no.: 63 82 35).

The controller can be operated via serial interface. Measuring results, preselections and the status of the
controller can be read at any time. Controlling via interface is only possible, if the remote operation mode
has been preselected at the controller. During remote operation the controller can be operated only via
interface, the keys at the controller have no function.

Setting of the interface

Setting of the interface parameters directly at the controller is described below. The factory set values are
underlined.
➨ 1200, 2400, 4800 or 9600 BAUD

➨ 7 data bits odd, 7 data bits even, 8 data bits none, 8 data bits odd, 8 data bits even

➨ no Handshake, XON/XOFF Handshake, CTS/RTS Handshake

➨ remote on, remote off

➨ Startbit = 1, Stopbit = 1

During remote operation all keys are without function.

Keep key

▲ pressed while switching on.

☞ The controller switches to the mode to configurate the inter-

face.

BAUD-Rate is displayed.
➨ Press key ▲ or ▼ to select the BAUD rate and confirm with

key START/STOP (”96”=9600, ”48”=4800, ”24”=2400,
”12”=1200).

Data bits are displayed.
➨ Press key ▲ or ▼ to select the data bits and confirm with

key START/STOP.

Parity is displayed.
➨ Press key ▲ or ▼ to select parity and confirm with key

START/STOP (”no” = no check of parity, ”O” = odd, ”E” =
even).

Handshake is displayed.
➨ Press key ▲ or ▼ to select handshake and confirm with key

START/STOP (”no” = no Handshake, ”HA” = RTS/CTS, ”SO”
= XON/XOFF).

Remote is displayed.
➨ Press key ▲ or ▼ to select local or remote operation and

confirm with key START/STOP (”no” = local operation, ”On”
= remote operation).

☞ The controller switches to normal operation mode (process

control not active).

+

(mains switch)

Set

b.96

Set

d.8

Set

P.no

Set

h.no

Set

r.no