Figure 3-5, Below-low analog triggering – ADLINK USB-1210 User Manual

Page 36

Advertising
background image

26

Operation & Calibration

trigger occurs when the analog trigger signal crosses above
(above high) or below (below low) the pre-defined voltage level.
The range of trigger level is the full-scale range of the selected
input channel and the resolution is 16-bit.

Below-Low Analog Triggering

In below-low analog triggering, as shown, the trigger signal is
generated when the input analog signal is less than the
Low_Threshold voltage. High_Threshold setting is not used in
this triggering situation.

Figure 3-5: Below-Low Analog Triggering

Above-High Analog Triggering

In above-high analog triggering, as shown, the trigger signal is
generated when the input analog signal exceeds the
High_Threshold voltage. Low_Threshold setting is not used in
this triggering situation