Gated trigger, Figure 3-12, Post trigger with retriggering – ADLINK USB-1210 User Manual

Page 40: Figure 3-13, 30 operation & calibration, Figure 3-12: post trigger with retriggering, Figure 3-13: gated trigger

Advertising
background image

30

Operation & Calibration

acquired after the first trigger signal, after which the USB-1210
awaits the next re-trigger signal (re-trigger signals occurring
before the N samples are completed will be ignored). When the
re-trigger signal occurs, another N samples are performed. The
process repeats until the specified number of re-trigger signals are
detected.

Figure 3-12: Post Trigger with Retriggering

Gated Trigger

Gated-trigger acquisition is indicated in applications where you
want to collect data when trigger events are set to level high/low,
and hold acquisition when trigger events are set to the opposite
level.

As shown, after the operation starts, the first scan of data is imme-
diately acquired when the trigger signal is deasserted and paused
when the trigger signal is asserted.

The remaining A/D conversions are not performed until the trigger
signal is deasserted again. The process repeats until the specified
amount of data is acquired.

Figure 3-13: Gated Trigger

Time

ƔOperation initiates

Trigger

Data

ƔTrigger event occurs
ƔAcquisition initiates

N samples

N samples

N samples

Time

ƔOperation initiates

Trigger

(Active High)

Data

ƔAcquisition begins

ƔAcquisition stops

ƔAcquisition begins ƔAcquisition stops

ƔAcquisition begins