ENMET PID-Guard User Manual

Page 11

Advertising
background image

PID-G

UARD

ENMET Corporation

9

4.2 Normal Display Mode

When the

PID-G

UARD

is installed as described in section 3, and in clean air, the

POWER

green LED is on, the display is lit

and the information on the display is measurement of hydrocarbon detected by the

PID-G

UARD

.

The red alarm and fault LEDs

are not lit.

To advance through displays of operational information press the

M

ENU

button.

Example sequence of operational information below:

Display Measurement of Hydrocarbon

Press

M

ENU

button

Display indicates Alarm 1 Set point

Press

M

ENU

button

Display indicates Alarm 2 Set point

Press

M

ENU

button

Display indicates Alarm 3 Set point

Press

M

ENU

button

Display indicates mA Span range
(Full Scale)

Press

M

ENU

button

Display returns to Hydrocarbon measurent

Operational Display Flow Chart

4.2.1 Alarm Conditions

PID-G

UARD

There are three alarm set points for CO. The factory settings of these alarm set points are shown in Table 2.

Table 2: Factory Alarm Set Points

Gas

Alarm 1

Alarm 2

Alarm 3

Hydrocarbon

5

10

50

These alarm set points can be changed within limits; see the maintenance section of this manual for the procedure.

If the gas concentration increases above that of the alarm set point, the associated red LED is lit, the associated relay changes
state, and the audio alarm is activated.

Pressing the

S

ELECT

button can temporally disable the Audio Alarm. The horn will be disabled for about five minutes. If a

second alarm condition occurs during this time the horn will re-activate. If the alarm condition(s) have ended during this time
the horn will not re-activate.

4.3 Hydrocarbon Sensor Response

Hydrocarbon (HC) sensor is designed to detect hydrocarbon gases and vapors with an ionization potential (IP) of 10.6

eV

or

less. Hydrocarbons with an IP of greater then 10.6

eV

will NOT be detected. Please see Appendix A for a list of common

gases and vapors and their respective IP rating.
Unless otherwise noted Isobutylene is used as a calibration and reference gas.

0 ppm

A1:

05

S

ELECT

S

ELECT

M

ENU

M

ENU

A2:

10

S

ELECT

M

ENU

A3:

20

S

ELECT

M

ENU

mA:

50

S

ELECT

M

ENU