Enter the maintenance menu. See Figure 7,

PID-G

UARD

Maintenance Menu flow chart.

1.

Press the

M

ENU

button until Span display.

2.

Press the

S

ELECT

button to perform a Span procedure.

The display will alternate between the calibration gas concentration (

Cal 20

) and a signal level (

PV

).

To Abort calibration press and Hold

M

ENU

button for 3 – 4 seconds, Abort? will appear, press

S

ELECT

button to return to

Span.
To change calibration gas level to be used, press and Hold

S

ELECT

button for 3 – 4 seconds, use menu button to change

digit and select button to move to next digit.

3.

Attach the associated calibration gas cylinder to the calibration adapter. See Figure 6 calibration adapter.

4.

Open the valve to apply the calibration gas to the sensor.
An auto detect sequence is initiated after 30 seconds, the

PID-G

UARD

will monitor the cal reading for stability.

5.

Watch for the signal level to stabilize. 1 – 4 minutes.

6.

Once the signal level has stabilized,

If the Span is successful, Cal OK appears momentarily, then will advance to Alarm1 menu.

If the sensor is outside of acceptable parameters, Bad Span is displayed.

If the sensor did not respond, an incompatible span gas was applied and the sensor did not respond at all,
Same mV is displayed then will return to Span.

If calibration is not successful, it is suggested that calibration be attempted again in 30-60 minutes.
If the sensor will not calibrate See Section 5.4.

7.

Remove the calibration gas.

8.

Calibration is complete.
Note: The instrument will return to operation mode in 3 – 5 seconds.

9.

Press the

M

ENU

button to advance to next desired menu

Example of Calibration Display:

S

ELECT

You can Press and H

OLD

the S

ELECT

button to change the

Calibration Gas Level
-Use the M

ENU

button to change digits

-Use the S

ELECT

button to move to next digit

N

OTE

:

To abort calibration or change calibration gas level.

You can Press and H

OLD

the M

ENU

button to abort Calibration

When Abort? appears press S

ELECT

button to advance to

Alarm1 menu

Abort?

S

ELECT

20

M

ENU

OR

OR

M

ENU

PV:

0

20

Apply Cal Gas until signal value becomes stable

(about 1 to 4 minutes) See Figure 7

When cal signal is stable PID-G

UARD

will automatically update:

If cal is good display will indicate OK or Same and advance to
Alarm1

If cal is not within preset “range” display will indicate Bad Sens
or Same mV The PID-G

UARD

will return to the Span Menu

To exit press M

ENU

button until Exit appears and press S

ELECT

Cal OK

Span

S

ELECT

S

ELECT

PV:

0

Same mV

Bad Sens

S

ELECT