ENMET PID-Guard User Manual

Page 18

Advertising
background image

PID-G

UARD

ENMET Corporation

16

5.4 Sensor Replacement

W

ARNING

:

Power must be removed from the

PID-G

UARD

before this or any internal procedure. Failure to do so may cause

damage to equipment, bodily injury or death.

Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in Section 6.0
of this manual. If you do not know the proper part number for your sensor, have the

PID-G

UARD

serial number available when

contacting your Distributor or

ENMET

Corporation Technical Support.

Remove, the 4 retaining screws from

PID-G

UARD

lid and 2 retaining screws form sensor manifold, see Figure 6

Remove, the sensor assembly from the PCB/sensor manifold, see Figure 8.

Insert, the new sensor assembly and replace sensor manifold and 2 retaining screws.

Replace, lid and the 4 retaining screws.

Re-supply power to the

PID-G

UARD

Figure 8: PID-G

UARD

Sensor Replacement

After the new sensor assembly has been installed, it is suggested to allow the sensor to stabilize for 3 – 4 hours.

A Factory calibration must be performed.

After entering the Maintenance menu, press and hold the

M

ENU

button for 2-4 seconds while viewing the Zero menu.

After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory
mode.

Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the
Zero and Span functions.

The Factory calibration sets a calibration window for future standard instrument calibrations.

PCB/Sensor
Manifold

Cover Inside View

Opened Upward

Attached to Base