3.0 Installation

The

PID-G

UARD

should be located near the pipe or tank containing the air to be monitored, and upstream from where the air is

being used. The

PID-G

UARD

must be installed such that it samples the compressed air before it reaches the users.

3.1 Mounting PID-G

UARD

Mount the

PID-G

UARD

instrument on an appropriate vertical surface, leaving room for lid to be opened, using the mounting

holes provided. Avoid areas with excessive vibration or temperature extremes. The holes in the bottom of the enclosure are
0.18 inch in diameter and form a 6.44

″ x 4.47″ rectangle. See Figure 3

It is recommended to use #8 drywall anchors and screws for mounting the

PID-G

UARD

to a drywall/sheetrock surface.

Dimensions are in inches.

Figure 3: Mounting PID-G

UARD

Right Side View

Cover Inside View

Opened Upward

Attached to Base