Pid-g, Maintenance menu flow chart 5ppm, Exit – ENMET PID-Guard User Manual

Page 17: Pv: 0 20, Cal ok same mv bad sens span, Alarm1, Alarm2, Alarm3, Ma span, Zero cal ok

Advertising
background image

PID-G

UARD

ENMET Corporation

15

Normal Gas Display

F

IGURE

7: PID-G

UARD

Maintenance Menu Flow Chart

5ppm

M

ENU

Press and H

OLD

the M

ENU

button for 2 – 4 seconds to enter the Maintenance Menus

Exit

S

ELECT

Press the S

ELECT

button to return to the Normal Gas Display.

See Section 5.2.1

M

ENU

Press the M

ENU

button to advance through Maintenance Menus cycle

OR

OR

M

ENU

PV:

0

20

Apply Cal Gas until signal value becomes stable
(about 1 to 4 minutes) See Figure 6

Press S

ELECT

to enter the cal signal:

If cal is good display will indicate OK or Same

If cal is not within preset “range” display will

indicate Bad Sens

Cal OK

Same mV

Bad Sens

Span

S

ELECT

S

ELECT

S

ELECT

You can Press and H

OLD

the S

ELECT

button to

change the Calibration Gas Level

See Section 5.2.3

Alarm1

M

ENU

Alarm2

M

ENU

Alarm3

S

ELECT

S

ELECT

S

ELECT

0020

0005

0005

0010

0010

To change Alarm set points:
Press Menu button until Alarm to be changed is displayed
Press Select button to display the set point

The M

ENU

button changes digit indicated by underscore

cursor
The S

ELECT

button locks underscored digit and moves to

next digit

If change is not within range display returns to first digit
If change is within range display moves to next menu

See Section 5.2.4

M

ENU

mA Span

M

ENU

S

ELECT

0050

To change mA Span set point:
Press Menu button until mA Span is displayed
Press Select button to display the set point

The M

ENU

button changes digit indicated by underscore cursor

The S

ELECT

button locks underscored digit and moves to next digit

See Section 5.2.5

To return to Normal Gas Display:
Press M

ENU

button until EXIT is displayed

Then press S

ELECT

button

OR

Zero

Cal OK

If the Zero signal is not within Preset Specs the

PID-G

UARD

will

display Bad ZERO

S

ELECT

M

ENU

Bad ZERO

If the Zero signal is within Preset Specs the

PID-G

UARD

will

display Cal OK,

See Section 5.2.2

PV:

0

S

ELECT

Press the S

ELECT

button to initiate Zero adjustment