3 circuit board features, Figure 2: pid-g, Circuit board features – ENMET PID-Guard User Manual

Page 6

Advertising
background image

PID-G

UARD

ENMET Corporation

4

2.3 Circuit Board Features

The Display Panel is attached by a cable and is released by unscrewing the 4 screws located in the corners. After releasing the
panel, it is swung upward, exposing the interior of the enclosure. The Circuit Board is mounted at the back surface of the
enclosure interior. Features are shown in Figure 2.

Feature

Description

Relay Terminals

This group of terminals is located on the Circuit Board.
For the contacts for each of three alarm relays, and for the contacts of a fault relay.
See Section 3.2.2

Output Terminals

For the 4-20 mA output.

PCB/Sensor Manifold

The PCB/Sensor manifold. The photoionization sensor located under this housing.

Figure 2: PID-G

UARD

Circuit Board Features

Relay Terminals
J14, J15, J16, J17

PCB/Sensor Manifold

Relays
K1, K2, K3, K4

Terminal J12

External Power

4-20mA Output