Hanna Instruments HI 120 User Manual

Page 18

Advertising
background image

18

CFM

CFM

• After the third calibration point is confirmed, immerse the pH electrode and the

temperature probe approximately 4 cm (1½”) into the fourth buffer solution
and stir gently. The temperature probe should be close to the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected

buffer, “CFM” tag blinks.

• Press CFM to confirm calibration.
• The calibrated value is then displayed on the

primary LCD and the secondary LCD will display the
fifth expected buffer value.

• After the fourth calibration point is confirmed, immerse the pH

electrode and the temperature probe approximately 4 cm (1½”) into
the fifth buffer solution and stir gently. The temperature probe should
be close to the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected

buffer, “CFM” tag blinks.

• Press CFM to confirm calibration.
• The instruments store the calibration values and

return to normal measurement mode.

FOUR, THREE or TWO-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL after the appropriate accepted calibration point. The instruments

will return to measurement mode and will memorize the calibration data.

ONE-POINT CALIBRATION

Two SETUP selectable options are available: “Pnt” and “OFFS”.
If the “Pnt” option is selected, the adjacent slopes will be reevaluated.
If the “OFFS” option is selected, an electrode offset correction is performed
keeping unchanged the existing slopes.
• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press CAL after the first calibration point was confirmed. The instruments

will memorize the one-point calibration data and will return to measure-
ment mode.

Notes: • Press MODE or / from the left keyboard to toggle between

pH buffer and temperature reading during calibration.

• Each time a buffer is confirmed, the new calibration parameters

replace the old calibration parameters of the corresponding buffer.
If current confirmed buffer has no correspondence in the existing
stored calibration and this is not full, the current buffer is added
to the existing stored calibration.