Hanna Instruments HI 120 User Manual

Page 43

Advertising
background image

43

CHR xx

Change the instrument range according with the parameter
value (xx):

• xx=00 pH range/0.001 resolution
• xx=01 pH range/0.01 resolution
• xx=03 mV range
• xx=04 Relative mV range

The meter will answer for these commands with:

<STX> <answer> <ETX>

where:

<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:

<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER

RAS

Causes the instrument to send a complete set of readings
in according with the current range:

• pH, mV and temperature reading on pH range.
• mV and temperature reading on mV range.
• Rel mV, absolute mV and temperature reading on

Rel mV range.

Note: The temperature is sent with two decimals and in Celsius degrees,

even if the set temperature unit was Fahrenheit degrees.

DA?

Requests the date.

TI?

Requests the time.

MDR

Requests the instrument model name and firmware code.

GLP

Requests the calibration data record.

PAR

Requests the setup parameters setting.

NSL

Requests the number of logged samples.

?ML

Requests the information about AutoLOG.

/MLPxxx

Selects the lot.

?DM

Downloads the selected AutoLOG.

LODPxxx Requests the xxx

th

record logged data.

LODPALL Requests all Log on demand.
Notes: • “Err8” is sent if the instrument is not in measurement mode.

• “Err7” is sent if a different range is requested during AutoLOGGING

mode (for CHR xx command only).

• “Err6” is sent if the requested range is not available.
• “Err4” is sent if the requested set parameter is not available.
• “Err3” is sent if the Log on demand is empty.
• Invalid commands will be ignored.