Hanna Instruments HI 120 User Manual

Page 28

Advertising
background image

28

If there is another item to be set (e.g. minutes), press MODE or /
from the left keyboard. The other item will start blinking.

Press the ARROW keys to change the displayed value or simply use the
numerical keys for all numerical parameters.
Press CFM to confirm or CAL to escape.
Press the ARROW keys to select the next/previous parameter.
Press 2nd then SET key, or simply SET from the left keyboard (HI 120 only)
to exit SETUP menu at any time. The instrument asks for printing a Setup
Report (HI 122 only). Press CFM to print the Setup Report or CAL to return
to measurement mode.
The following table lists the SETUP parameters, their valid values range
and the factory settings (default):

PrEF

Command Prefix

0 to 47

16

In Id Instrument ID

0000 to 9999

0000

bEEP

Beep Status

ON/OFF

OFF

bAUd Baud Rate

600; 1200; 2400; 4800; 9600

2400

LAnG

Printing Language

EnG, ItA, ESP, FrA

English USA

Printer Status

ON/OFF

OFF

Custom C2 Custom Buffer 2

-2.00 to 16.00 pH

no

Custom C1 Custom Buffer 1

-2.00 to 16.00 pH

no

Date

Date (yyyy.mm.dd)

2000.01.01 to 2099.12.31

2004.01.01

CAL DUE Alarm Time Out

OFF or 1 to 7 days

OFF

Time

Time (hh:mm)

00:00 to 23:59

00:00

Strt

AutoLOG Start Condition

See Time/Date or “btn”

“btn”

End

AutoLOG End Condition

“dur”, “SAMP”, “rdG”

“dur”

Interval AutoLOG Interval

5 s to 180 min

5 s

AL.HI Alarm High Limit

pH/mV/RelmV ranges

no

AL.LO Alarm Low Limit

pH/mV/RelmV ranges

no

Item

Description

Valid values

Default

1Pnt

One-point cal. behaviour

Pnt or OFFS

Pnt

tEMP

Temperature Unit

ºC or ºF

ºC