Hanna Instruments HI 120 User Manual

Page 47

Advertising
background image

47

pH CLEANING PROCEDURE

• General

Soak in Hanna HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour.

• Protein

Soak in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.

• Inorganic

Soak in Hanna HI 7074 Inorganic Cleaning
Solution for 15 minutes.

• Oil/grease

Rinse with Hanna HI 7077 or HI 8077 Oil and
Fat Cleaning Solution.

IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water, refill the reference chamber with
fresh electrolyte (not necessary for gel-filled electrodes) and soak the electrode
in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking
measurements.

Advertising
This manual is related to the following products: