36

GLP Rel mV

Instr ID 0002
Date 2004/10/15
Time 14:20:05
Off.RelmV 0.3

A complete set of information based on the measured, set or recorded data
can be printed.
Data can be printed on demand (for current reading in measurement
mode, GLP and SETUP modes) by pressing 2nd then Print key, or simply Print
from the left keyboard, or automatic (for AutoLOG and Log on demand modes).
Automatic data printing is possible in AutoLOG mode only if the printer is
ON and the AutoLOG interval is greater than 30 seconds.
Note: To cancel printing, press 2nd then Print key again or simply Print

from the left keyboard.

When in pH measurement mode, the
printout provides the following infor-
mation:

Note: For Relative mV measurement mode, the Rel. mV value will be printed.
When in GLP mode, the printout provides the following information:
• For pH range:

• For Rel mV range:

Date 2004/10/15
Time 14:24:55
pH 7.00
mV 0.0
°C 25.0

PRINTING (HI 122 only)

GLP pH

Instr ID 0002
Date 2004/10/15
Time 15:30:05
Cal Time Out OFF
Offset

0.6mV

Slope 99.7%
Cal Buffers
pH 7.01
pH 4.01
pH 10.01